All published articles of this journal are available on ScienceDirect.

RESEARCH ARTICLE

Efficacy and Safety of Cataract Extraction with Negative Power Intraocular Lens Implantation§

The Open Ophthalmology Journal 15 Feb 2008 RESEARCH ARTICLE DOI: 10.2174/1874364100802010015

Abstract

Purpose:

To evaluate the visual and refractive outcomes, lens power calculation accuracy, and safety of negative power intraocular lenses (IOLs) implanted in highly myopic eyes at the time of cataract surgery.

Design:

Interventional case series.

Methods:

Sixteen consecutive highly myopic eyes implanted with IOLs from –1 D to –6 D were identified. IOL power; preoperative and postoperative best-corrected visual acuity (BCVA); postoperative uncorrected visual acuity (UCVA); preoperative, intended, and achieved spherical equivalent (SE) refractive errors; and operative complications were recorded.

Results:

Median UCVA improved from finger counting to 20/50-2. Median BCVA improved from 20/125-1 to 20/30+1. Mean axial length was 32.65 mm. The mean SE refractive error was –22.19 ± 5.4 D before surgery and -0.28 ± 1.4 D after surgery. The difference between the mean intended and mean achieved SE refractive errors was +1.16 D for the SRK/T, +1.2 D for the Holladay 1, and +1.60 D for the Hoffer Q formulas. Only 5 (33.3%) of 15 eyes in which postoperative measurements were possible were within 1 D of the intended SE postoperative refraction. Postoperative complications included a mildly hyperopic postoperative refractive error (+1.75 D) in one eye necessitating an IOL exchange and posterior capsule opacification in most eyes. There were no retinal detachments.

Conclusions:

The SRK/T formula had the greatest accuracy and predictability when immersion A-scan ultrasonography was used to measure axial length. The mean achieved postoperative refractive error was +1.16 D more hyperopic than predicted by this formula. We recommend targeting highly myopic eyes for –1.5 D using the SRK/T formula if a negative power IOL is calculated and emmetropia or mild residual myopia is the desired postoperative result.

Fulltext HTML PDF