Comparison of Clinical Results, Contrast Sensitivity and Optical Quality Following the Implantation of CT LUCIA 611P and TECNIS-1 ZCB00 MonofocalIOLs -12 Month Outcomes
RESEARCH ARTICLE

Comparison of Clinical Results, Contrast Sensitivity and Optical Quality Following the Implantation of CT LUCIA 611P and TECNIS-1 ZCB00 MonofocalIOLs -12 Month Outcomes

The Open Ophthalmology Journal 01 Nov 2022 RESEARCH ARTICLE DOI: 10.2174/18743641-v16-e2208191

Abstract

Purpose:

To compare the clinical performance, contrast sensitivity and optical quality, following implantation of CT LUCIA 611 P and TECNIS-1 monofocal IOLs following cataract surgery.

Design-Prospective, interventional, non-randomized comparative study.

Setting- Nethradhama Super Speciality Eye Hospital, Bangalore, India.

Methods:

Eligible patients, undergoing phacoemulsification received implantation with CT LUCIA 611P or TECNIS-One piece (TECNIS-1) monofocal IOLs.

Results:

Hundred eyes from 100 patients were sequentially divided into CT LUCIA and TECNIS-1 groups, with 50 eyes in each group.Intra-operatively, the mean unfolding time was significantly longer (35.16 ± 10.50 sec) in the TECNIS-1, compared to the CT LUCIA group (12.93 ± 3.80 sec), p= 0.00.At 12 months, 80% (40) eyes in the CT LUCIA and 76%(38) eyes in the TECNIS-1 group had cumulative UDVA of 20/20 or better. No significant differences were found between the mean values of post-op UDVA, CDVA, contrast sensitivity (all spatial frequencies), Objective Scatter Index (OSI), and Modular Transfer Function (MTF) between both groups. A significantly higher value of internal coma and SA for the Tecnis-1 IOL group was noted (p<0.05). However, there was no significant difference between the total HOA, coma and SA for both the groups. Six eyes in the TECNIS-1 group had intra-operative adhesions of the haptics with optic / haptic, requiring additional manipulation.

Conclusion:

At one year, both monofocal IOLs delivered comparable clinical outcomes.However, CT LUCIA 611P IOL had significantly less internal coma and SA, unfolding time and smoother IOL insertion without any issues due to poor loading.

Keywords: CT LUCIA, TECNIS-1, Monofocal, IOL, Visual performance, Optical quality.

1. INTRODUCTION

While there is no ideal monofocal intraocular lens (IOL) available as yet, significant improvements in the field of monofocal IOL technology have occurred in the past few decades to achieve an IOL design as close to perfection. A novel IOL design must offer excellent optical performance without unwanted side effects, should be made using a high-purity material to reduce the incidence of glistening, has good capsular and uveal biocompatibility, possess optics that account for eyes requiring aspheric, spheric or neutral corrections and feature a design that fits through small corneal incisions to allow for optimal centration in the capsular bag [1-3].Apart from these, the lens should include square edge technology to prevent posterior capsule opacification (PCO) [4, 5].

A new monofocal IOL is thus expected to fulfil all these properties. In 2016, a hydrophobic acrylic, heparin-coated, fully pre-loaded monofocal IOL,CT LUCIA 611P IOL(Carl Zeiss Meditec, Jena, Germany) was introduced as a successor of CT LUCIA 601 with certain modifications in the design of its optic-haptic junction for better capsular bag stability [6].

In the present study, we compared the 12-month post-operative visual outcomes, refractive stability, aberrations and rate of posterior capsule opacification (PCO) formation following implantation of CT LUCIA 611 IOL with TECNIS-1 ZCB00 IOL; the safety, efficacy and optical quality of which has already been established in various clinical studies [7-9].We also wanted to verify, if the heparin coating on the CT LUCIA 611 IOL, lead to an increased scatter or degradation of optical quality. Hence, Objective Scatter Index (OSI), Modular Transfer Function (MTF)and contrast sensitivity were also compared between the two IOLs.

2. MATERIALS AND METHODS

This prospective study was approved by the institutional ethics committee of Nethradhama Eye Hospital, Bangalore, and adhered to the tenets of the declaration of Helsinki. All patients provided written informed consent. A total of 100 eyes from 100 patients were included in the study and divided into two groups: CT LUCIA and TECNIS-1, with 50 eyes in each group.The first group of 50 patients who met the enrolment criteria received CT LUCIA, while the next group of 50 patients received the TECNIS-1 monofocal IOL.

Inclusion criteria were healthy eyes besides senile cataract NC 1-2 (LOCS III grading); corneal astigmatism equal to or less than 0.75 dioptres (D); IOL powers between +10.00 D and +32.00 D, in the capsular bag IOL implantation. Exclusion criteria were eyes with irregular astigmatism, corneal dystrophy, pupillary abnormalities, history of glaucoma or intraocular inflammation, macular disease or retinopathy, neuro-ophthalmic diseases, and intraoperative or postoperative complications.

Before the surgery, all patients underwent complete ophthalmologic examination including manifest refraction, slit lamp biomicroscopy, and noncontact tonometry (Tomey NCT, NishiKu, Nagoya, Japan), macular OCT (Optovue, Fremont, USA) and dilated fundus examination.

Biometric assessments were performed using the swept source OCT based optical biometer, IOL Master-700, using Barrett’s Total Keratometry (TK) formula [10]. All eyes were targeted at emmetropia. The optimized A- constants of 119.9 and 119.15 were used for the CT LUCIA 611 P and TECNIS-1 IOLs respectively.

2.1. Description of Study IOLs

Table 1 shows the comparison of specifications of CT LUCIA 611Pand TECNIS-1 monofocal IOLs.

The CT LUCIA 611P (Carl Zeiss Meditec, Jena, Germany) is a fully preloaded hydrophobic, acrylic, single-piece, heparin-coated IOL with an overall diameter of 13 mm and an optic diameter of 6 mm. The lens is available in clear UV-blocking (611P) and with blue light filtering (yellow tinted as 611Y) in a range of 4.00–34.00 D in 0.50 D increments. It is completely preloaded in the Zeiss Blueject injector to facilitate a fast and easy lens preparation. The 611P is equipped with a 360° square edge design on the entire IOL, including the optic, the haptics and the optic–haptic transition to prevent PCO. In addition, the haptics is step-vaulted to translate the optic posteriorly for direct contact with the capsular bag. The optic of the lens has a special patented aspheric design, called ZO optic, to compensate for the range of aberrations that arise from different corneal shapes and lens misalignments.

The Tecnis 1-piece IOL(Johnson & Johnson, New Jersey, USA) has a 6.0 mm biconvex optic, an overall diameter of 13.0 mm, and an anterior aspheric surface. The optic has a continuous 360-degree square frosted edge. The C-shaped haptics are offset from the optic for 3-point fixation. The optic and haptics are of an ultraviolet light–filtering hydrophobic acrylic material.

Table 1.
Comparison of specifications of CT LUCIA 611Pand TECNIS-1 monofocal IOLs.
Feature CT Lucia 611 P Tecnis-1
Optic design Monofocal aspheric Monofocal
Shape Biconvex, anterior aspheric surface, square optic edge Biconvex, anterior aspheric surface, square optic edge
Material Hydrophobic acrylic with heparin coated surface and blue light filter UV blocking hydrophobic acrylic
Refractive index 1.49 1.47
Optic diameter 6.0 mm 6.0 mm
Total diameter 13.0 mm 13.0 mm
Haptic Step vaulted Haptic offset from optic
Lens design Single piece Single piece
Incision size 2.2-2.6mm 2.2-2.4mm
Company labelled A constant 119.9 119.3 (Optical biometry)
118.8 (Ultrasound biometry- Contact)
Dioptre range +4.0 to +34.0,
0.5 D increments
+5.0 to +34.0,
0.5 D increments
ACD 6.14
Abbe number 50 55
Implantation in Capsular bag Capsular bag

2.2. Surgical Procedure

All surgeries were performed by a single experienced surgeon (S.G.), using a standard phacoemulsification technique under topical anaesthesia. All surgeries in both groups were performed from the temporal site, through a clear corneal incision of 2.8 mm size, using the Centurion Precision system (Alcon). A standard capsulorhexis of between 5.0 -5.5 mm was aimed and a direct chop technique was used for nuclear deployment. In the CT LUCIA group, the IOL injection was facilitated through the fully pre-loaded Zeiss Blueject injector, while in the TECNIS-1 group, the UNFOLDER Platinum 1 Series Screw-Style Inserter (Johnson & Johnson) was used to inject the TECNIS-1 IOL through a 2.8 mm temporal clear corneal incision. IOL insertion was done under cover of OVD.

Intra-operative unfolding time was recorded by an independent observer (S.B.) using a stopwatch. Time was documented beginning from when the trailing haptic folded on the optic released from the injector, and stopping when both the IOL haptics completely unfolded inside the capsular bag. Note of any intra-operative complications or difficulty with IOL injection was made.

Post-operative topical therapy included topical prednisolone (1%, Pred Forte, Allergan), moxifloxacin (0.5%, Vigamox, Alcon), and nepafenac (0.1%, Nevanac, Alcon).

Post-operative follow-up examinations were performed at 1 day, 2 weeks, 6 weeks, 6 months, and 12 months. The following tests were performed at all post operative visits from the first week onwards: Slit lamp examination for corneal clarity, anterior chamber inflammation, IOL position, posterior capsule opacification (PCO), measurement of uniocular uncorrected (UDVA) and corrected distance visual acuity (CDVA) using ETDRS charts (Precision Vision, La Sella, IL, USA), uniocular uncorrected (UNVA), near corrected near visual acuity (CNVA); and mesopic contrast sensitivity (F.A.C.T. Stereo Optical Co. Inc., Chicago) with distance correction. Internal and total aberrations were measured using ray tracing aberrometry (I-trace, Hoya, Japan). A double pass system (HD Analyzer,Visiometrics, Spain) was used to evaluate the OSI and MTF cut-offs. This diagnostic tool measures both the scattered light and higher order aberrations, thus enabling objective assessment of optical quality. All measurements were recorded 5 minutes after instillation of a lubricant eye drop [Lacryl Ultra (polyethylene glycol 0.4%, propylene glycol 0.3%), ENTOD pharmaceuticals] as the dry-eye disease can degrade the quality of retinal image, resulting in higher OSI values.

2.3. Statistical Analysis

SPSS software for Windows version 17.0.0 (IBM Corp., Armonk, NY) was used for statistical analysis. All values were expressed as mean ± standard deviation (SD). Data was checked for normality before subjecting to analysis. If the data was normally distributed, then the Independent sample t-test was used for intergroup comparison and paired t-test was used for intra-group comparison of means. If the data distribution was not normal, Wilcoxon signed-rank test was used. A p-value of 0.05 or less was considered statistically significant. Outcome analysis was performed according to the Standard Graphs for Reporting Refractive Outcomes Intraocular Lens-Based Refractive Surgery [11].

3. RESULTS

Table 2 shows the pre-operative characteristics and demographic data of both the study groups.

Table 2.
Pre-operative characteristics and demographic data of both the study groups.
Parameter
(mean ±SD)
Range [min, max]
CT Lucia 611 P
(n=50)
Tecnis-1
(n=50)
p-value
Pre-operative
Age(years) 67.81± 8.9
[45.79]
67.79± 5.7
[60.80]
0.09
AL(mm) 23.80 ± 1.21
[22.13, 27.95]
23.76 ± 1.13
[21.9, 26.5]
0.86
K1(D) 43.64 ± 1.80
[49.92, 47.94]
43.82 ± 1.20
[41.62, 46.45]
0.56
K2(D) 44.16 ± 1.83
[41.39, 48.68]
44.28 ± 1.18
[42.23, 46.7]
0.69
Ast (D) 0.48± 0.26
[0.13, 0.75]
0.46± 0.20
[0, 0.75]
0.56
ACD(mm) 3.22 ± 0.45
[2.22, 4.72]
3.16 ± 0.41
[2.1, 4.42]
0.45
WTW(mm) 12.01 ± 0.45
[11, 13.1]
11.98 ± 0.41
[11.2, 12.9]
0.78
IOL power(D) 20.47 ± 2.93
[9.5, 24]
20.00 ± 3.21
[11.5, 24.5]
0.44
Intra-operative
Un-folding time (secs) 12.93± 3.80
[4.53, 26.25]
35.16 ± 10.50
[14.24, 56.8]
<0.05
Abbreviations: AL: axial length, mm: millimeter, K: keratometry, D: dioptre, Ast: astigmatism, ACD: anterior chamber depth, WTW: white to white, min: minimum, max: maximum
Fig. (1). Anterior chamber reaction for both the study IOLs on day 1 and 2 weeks post-operatively.

Both groups were matched in terms of mean age, axial length, K1, K2, corneal astigmatism, anterior chamber depth (ACD), white to white (WTW) diameter and IOL powers implanted(p-values >0.05, for all parameters).

3.1. Post-operative AC Reaction

Both groups showed comparable results in terms of postoperative anterior chamber reaction on both days 1 and 2 weeks. No eye in either group had excessive inflammation of more than 2+ cells on post-op day 1 and more than 1+ cells at 2 weeks, (Fig. 1). At 6 weeks post-op, anterior chamber in all eyes were quiet and did not show any evidence of cells or flare.

3.2. Visual outcomes

Thirty six percent (18) eyes in the CT LUCIA group had post-operative CDVA one line or better than the post-operative UDVA, whereas this percentage for the Tecnis group was 44% (22 eyes) (Table 3) (Figs. 2 and 3).

Fig. (2). Histogram showing UDVA and CDVA for both the study IOLs at 12 months post-operatively.
Fig. (3). Histogram showing the change in Snellen’s lines of CDVA for both the study IOLs at 12 months post-operatively.

3.3. Refractive outcomes

Ninety four percent (n=47) eyes in both groups had SE accuracy within ±0.50 D and all eyes had SE accuracy within ±1.00 D, (Fig. 4).

Post-operative astigmatism was within ±1.00 D in both the groups, (Fig. 5).

3.4. Contrast Sensitivity

The results of the mean contrast sensitivity values for both the monofocal IOL groups, measured after correction, using the F.A.C.T. chart in photopic conditions, 12 months post-operatively did not show a significant difference for any of the spatial frequencies evaluated (p>0.05, for all spatial frequencies) (Table 2) (Fig. 6).

Fig. (4). Histogram showing the accuracy to the intended spherical equivalent refraction for CT LUCIA 611 P and TECNIS-1 eyes at 12 months post-operatively.
Fig. (5). Histogram showing change in refractive astigmatism for both the study IOLs at 12 months post-operatively.
Fig. (6). Mean contrast sensitivity (photopic) values for both the monofocal IOL groups at 12 months post-operatively.

3.5. OSI and MTF

OSI and MTF remained stable through 12 months, with no significant change compared to the 1-month values (One -way ANOVA, p > 0.05 in both the groups), Table 3. At 12 months, the mean OSI values of the CT LUCIA group was 0.90± 0.45 and in the TECNIS-1 group was0.78± 0.45,which was not statistically significantly different. Similarly, there was no significant difference between the mean MTF for the CT LUCIA group (0.47± 0.14)and the TECNIS-1 group (0.49± 0.12) at the last visit (Fig. 7).

3.6. Higher Order & Internal Aberrations

Higher order aberrations evaluated at 4 mm scan size using I-trace, showed a significantly higher value of internal coma (p=0.05) and SA for the TECNIS-1 IOL group (p=0.01). However, there was no significant difference between the mean RMS of internal HOA for both groups. For the total HOA, coma and SA, all values between both the groups were comparable (Table 4).

Fig. (7). Objective Scatter Index (OSI) and Modular Transfer Function (MTF), for both the study IOLs, at 12 months post-operatively.
Table 3.
Comparison of visual outcomes, OSI and MTF at 1 month, 6 months and 12 months post-operatively.
Parameters
(Mean± SD)
Range [min, max]
1 month 6 months 1 year p-value

CT Lucia 611 P
UDVA(LogMAR) 0.028± 0.08
[-0.1, 0.2]
0.022± 0.08
[-0.1, 0.2]
0.022± 0.08
[0, 0.2]
0.99
CDVA(LogMAR) -0.020± 0.06
[-0.1, 0.2]
-0.022± 0.06
[-0.1, 0.2]
-0.022± 0.06
[-0.1, 0.2]
0.98
Sph (D) -0.065± 0.38
[-0.5, 0.5]
-0.06± 0.35
[-0.5, 0.5]
-0.06± 0.35
[-0.5, 0.5]
0.96
Cyl (D) -0.11± 0.36
[-0.5, 0.75]
-0.11± 0.35
[-0.5, 0.75]
-0.11± 0.35
[-0.5, 0.75]
0.99
SE(D) -0.12± 0.31
[-0.75, 0.75]
-0.11± 0.29
[-0.75, 0.75]
-0.11± 0.29
[-0.25, 0.5]
0.97
ACD 5.05± 0.49
[3.33, 5.75]
5.01± 0.44
[3.66, 5.7]
4.98± 0.50
[3.25, 5.68]
0.71
HD Analyzer
OSI 0.92± 0.46
[0.3, 2.3]
0.91± 0.45
[0.3, 2.3]
0.90± 0.45
[0.3, 2.3]
0.99
MTF 0.46± 0.13
[0.163, 0.7]
0.47± 0.13
[0.163, 0.7]
0.47± 0.14
[0.163, 0.7]
0.89

Tecnis-1
UDVA(LogMAR) 0.034± 0.08
[-0.1, 0.3]
0.026± 0.08
[-0.1, 0.3]
0.026± 0.08
[-0.1, 0.3]
0.83
CDVA(LogMAR) -0.028± 0.04
[-0.1, 0.1]
-0.028± 0.04
[-0.1, 0.1]
-0.028± 0.04
[-0.1, 0.1]
0.97
Sph (D) -0.02± 0.16
[-0.5, 0.5]
-0.03± 0.32
[-0.5, 0.5]
-0.03± 0.32
[-0.5, 0.5]
0.93
Cyl (D) -0.12± 0.44
[-1, 0.75]
-0.12± 0.43
[-1, 0.75]
-0.12± 0.43
[-1, 0.5]
0.99
SE(D) -0.11± 0.29
[-0.75, 0.75]
-0.12± 0.27
[-0.75, 0.75]
-0.12± 0.27
[-0.75, 0.75]
0.98
ACD 5.03± 0.48
[3.33, 5.75]
4.87± 0.41
[4, 5.8]
4.83± 0.43
[4, 5.8]
0.58
HD Analyzer
OSI 0.80± 0.46
[0.3, 2.3]
0.79± 0.46
[0.3, 2.4]
0.78± 0.45
[0.3, 2.3]
0.98
MTF 0.47± 0.17
[0.189, 0.7]
0.48± 0.12
[0.189, 0.7]
0.49± 0.12
[0.189, 0.7]
0.72
Abbreviations: UDVA: uncorrected distance visual acuity, CDVA: corrected distance visual acuity, D: dioptre, Sph: sphere, Cyl: cylinder, SE: spherical equivalent, ACD: anterior chamber depth, OSI: objective scatter index, MTF: modulation transfer function, FACT: functional acuity contrast test, cpd: cycles per degree; min: minimum, max: maximum
Table 4.
Comparison of Internal and Total HOA values at 12 months post-operatively.
Parameters (µ)
(mean ± SD)
Range [min, max]
CT Lucia 611 P
(n=50)
Tecnis-1
(n=50)
p-value
Internal
HO Total 0.161± 0.072
[0.019, 0.352]
0.180± 0.191
[0.034, 1.95]
0.51
Coma 0.045± 0.043
[0.01, 0.2]
0.060± 0.050
[0.012, 0.2]
0.05
SA -0.050± 0.063
[[-0.14, 0.045]
-0.080± 0.054
[-0.132, 0.04]
0.01
Total
HO Total 0.203± 0.115
[0.05, 0.43]
0.168± 0.080
[0.054, 0.447]
0.08
Coma 0.043± 0.100
[-0.126, 0.242]
0.071± 0.049
[0.011, 0.362]
0.10
SA 0.029± 0.047
[-0.06, 0.09]
0.026± 0.021
[-0.029, 0.101]
0.75
Abbreviations: HO: higher order, SA: spherical aberration, min: minimum, max: maximum

3.7. Intra-operative unfolding time and adverse events

The mean intra-operative unfolding time was significantly higher (35.16 ± 10.50) in the TECNIS-1 group, compared to the CT LUCIA group (12.93 ± 3.80), p= 0.00 Six eyes in the TECNIS-1 group had intra-operative adhesions of the haptics with optic / opposite haptic, requiring additional manipulation to release the same with a second instrument. However, no eye in either group had an occurrence of trapped trailing haptic, overriding of the plunger over the optic, trauma to the optic edge or breakage of haptics. All procedures in both groups were otherwise uneventful. No post-operative complications such as cystoid macular oedema, post-op uveitis, secondary glaucoma or posterior capsule opacification requiring YAG- Capsulotomy were noted in any of the eyes of either group. No IOL in either of the study group required to exchange or explantation due to any reason.

4. DISCUSSION

Recently, Borkenstein et al. evaluated one-year clinical outcomes of a novel designed hydrophobic, acrylic, monofocal, and fully preloaded CT LUCIA 611P intraocular lens.6The optic–haptic junction (Achilles Heel) of the intraocular lens (studied with scanning electron microscopy)was found to be thicker, wider and stiffer than the predecessor (601), with a 360° square edge technology incorporated on it. The post-operative SE was within ±0.50 D in 91.7% and within ±0.75 D in 96.9% of cases. From the surgeons’ perspective, the wider and thicker optic–haptic transition, resulted in increased stiffness, enabling improved centration, refractive predictability, and stability of the IOL [6].In another study, they demonstrated that the CT LUCIA 611 P IOL resulted in good surgical performance, excellent refractive stability, and predictable outcomes in patients with PXF syndrome, phacodonesis and small pupils, where the stability of the capsular bag is already compromised [12].

In the present study, in terms of stability, both CT LUCIA 611 P and TECNIS-1 IOL showed similar stability in the capsular bag. The post-operative ACD remained stable, and no significant myopic or hyperopic shift was observed at 12 months, compared to 1 month in either of the groups. The newly designed thicker and stiffer optic–haptic junction thus appeared to improve the stability of the CT LUCIA IOL in the capsular bag, preventing any significant refractive changes in the long term.

Optical quality evaluated with HD Analyzer did not show any significant difference between the mean OSI and MTF values between both the lenses, at any post-operative time point. No eye in either group showed any evidence of PCO at 12 months due to the continuous 360 degrees square edge, effectively preventing LEC migration [7]. It may be inferred, that in the absence of PCO, the heparin-coated surface of the CT LUCIA IOL, did not result in increased scatter, and subsequent increase in the OSI values. The high and comparable MTF values of both the IOLs may be indicative of an absence of glistening due to the excellent quality of the material used [6, 13].

The perceived benefits of pre-loaded injectable IOL delivery systems include the elimination of the IOL injector loading variability seen in non-preloaded systems, avoidance of potential loading error, and reduced surgical time [14].Intraoperative problems with acrylic intraocular lens (IOL) insertion and post-operative implications due to this have been previously reported [15].Improper unfolding caused by one of the haptics sticking to the optic is known to occur due to inadequate OVD in the cartridge or rarely by the incorrect loading of the IOL [16].In a study evaluating the delivery characteristics of the AcrySof IQ SN60WS intraocular lens (IOL) injected via a preloaded AcrySert delivery system, 47 of the 85 eyes (55%) required additional rotational manipulation, management of trapped trailing haptic, haptic-optic adhesion, overriding of the plunger over the optic, and trauma to the optic edge [17].On the other hand, the fully preloaded feature of the CT LUCIA611 IOL was clearly seen as an advantage in the present study, as all implantations with CT LUCIA 611 IOL were smooth and problem free.

It is known that the mechanical properties of most polymers, including acrylics, are affected by the temperature, and the glass-transition temperature (Tg) of the polymer determines the ideal temperature for optimal unfolding within the eye [18].Chung et al.compared the characteristics of 5 different pre-loaded and non-preloaded intraocular lens (IOL) delivery, systems and found that the average time for non-preloaded systems was comparatively higher than the pre-loaded ones. MX60 had the highest IOL unfolding time in the capsular bag, due to its high “Tg” [19]. The glass transition temperature of TECNIS-1 IOL being comparatively higher (13.8° C) than CT LUCIA (11-12° C), this may explain the significantly shorter unfolding time of the latter, in the present study. In addition, the Heparin Surface Modification (HSM) provided on the CT LUCIA 1OL may also play a role in its faster unfolding.

In terms of the post-operative higher order aberrations, the internal coma and SA (i.e. arising from the internal optics) were found to be significantly higher in the TECNIS-1 group. This could suggest a higher degree of IOL tilt inside the capsular bag in this group. The internal spherical aberrations are expected to be higher with the TECNIS-1 IOL, due to the relatively higher negative SA (-0.27 μm) compared to the CT LUCIA 611P IOL (-0.12μm) [6, 7].It has been proposed that the CT LUCIA 611P IOL is equipped with ZO(600) aspheric concept optic, which means that the power of the IOL is higher in the centre and varies towards the periphery. This results in a flatter lens surface at an intermediate distance from the lens axis and a steepening at the peripheral region of the lens, which may potentially compensate for different corneal shapes and reduce the incidence of higher-order aberrations (HOAs). This should also make the lens less sensitive to decentration and tilt [6]. The ZO aspheric optic, combined with the optic-haptic modification, maythus explain the significantly lesser internal coma seen in the CT LUCIA group, suggestive of better centration in the capsular bag.

TheTECNIS-1IOL with more negative SA (IOL power decreases from centre to periphery) is designed to compensate for corneal aberrations and improve contrast sensitivity [7]. However, this doesn’t appear to have conveyed any added benefit, as the total SA was similar in both groups. As postulated earlier, lenses with higher negative asphericity, if not well aligned, may show increased HOAs such as coma, potentially degrading the visual quality [19-21].In our study, although the Tecnis-1 group had significantly higher internal coma, it did not seem to clinically affect the visual quality, as finally it is the total (whole eye) aberrations that matter.In the study, we did not evaluate corneal aberrations pre-operatively, as the main aim was to determine post-op internal aberrations (indirectly suggesting IOL position) and total aberrations (sum of corneal and internal aberrations) which determine the overall visual quality. It may be possible that the internal coma was compensated by the corneal aberrations in this group, and did not lead to visual degradation.

Heparin Surface Modification (HSM) of hydrophobic acrylic IOLs was suggested as a means to reduce post-operative inflammation [22].However, our study, did not find any clinically significant difference in the post-operative inflammation between both the groups. This may be due to the reason that we included only eyes with no other ocular co-morbidity besides age related cataracts such as pseudo exfoliation, chronic uveitis etc.The HSM of the CTLUCIA 611 IOL, however, did not seem to affect the optical quality,which was evident from comparable values of OSI, MTF and contrast sensitivity between the two groups at 12 months post-operatively.

CONCLUSION

In conclusion, both monofocal IOLs delivered excellent and comparable outcomes in terms of visual and refractive results, long-term stability, total aberrations and optical quality. However, CT LUCIA 611P IOL had significantly less internal coma and SA, unfolding time and uneventful IOL insertions. The Blueject injector appears to meet the expectations of providing a predictable means of IOL delivery.This may be considered a significant advantage, as increased manipulation of foldable IOLs within their injectors may render the IOLs more prone to damage and increase the risk of inadvertent intra-operative complications. Further studies may be beneficial to understand its long-term safety, efficacy, optical quality, PCO behaviour and capsular bag stability, in comparison with other concurrent monofocal IOL technologies.

LIST OF ABBREVIATIONS

OSI = Objective Scatter Index
MTF = Modular Transfer Function
PCO = Posterior Capsule Opacification

AUTHORS' CONTRIBUTIONS

Sheetal Brar – Data analysis, Manuscript writing, Statistical analysis, Review of literature.

Sri Ganesh- Concept and Design, Critical revision of manuscript.

Smith Snehal Sute- Data collection, Statistical analysis, Review of literature.

Swati Chidre- Data collection.

ETHICS APPROVAL AND CONSENT TO PARTICIPATE

This prospective study was approved by the institutional ethics committee of Nethradhama Eye Hospital, Bangalore.

HUMAN AND ANIMAL RIGHTS

No animals were used for studies that are the basis of this research. All the humans were used in accordance with the Helsinki Declaration of 1975, as revised in 2013 (http://ethics.iit.edu/ecodes/node/3931).

CONSENT FOR PUBLICATION

All patients provided written informed consent.

STANDARDS OF REPORTING

STROBE guidelines were followed

AVAILABILITY OF DATA AND MATERIALS

Supplementary material is available on the Publisher’s website.

FUNDING

None.

CONFLICT OF INTERESTS

The authors declare no conflict of interest, financial or otherwise.

ACKNOWLEDGEMENTS

Declared none.

REFERENCES

1
Leaming DV. Practice styles and preferences of ASCRS members—2003 survey. J Cataract Refract Surg 2004; 30(4): 892-900.
2
Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg 2010; 36(8): 1398-420.
3
Packer M, Rajan M, Ligabue E, Heiner P. Clinical properties of a novel, glistening-free, single-piece, hydrophobic acrylic IOL. Clin Ophthalmol 2014; 8: 421-7.
4
Dewey S. Posterior capsule opacification. Curr Opin Ophthalmol 2006; 17(1): 45-53.
5
Nanu RV, Ungureanu E, Istrate SL, et al. An overview of the influence and design of biomaterial of the intraocular implant of the posterior capsule opacification. Rom J Ophthalmol 2018; 62(3): 188-93.
6
Borkenstein AF, Borkenstein EM. Long-term clinical results and scanning electron microscopic analysis of the aspheric, hydrophobic, acrylic intraocular lens CT LUCIA 611P(Y). Clin Ophthalmol 2018; 12: 1219-27.
7
Wahba S, Riad RF, Morkos FF, Hassouna AK, Roshdy MM. Visual performance of the tecnis one-piece lens ZCB00. Clin Ophthalmol 2011; 5: 1803-8.
8
Jafarinasab MR, Feizi S, Baghi AR, Ziaie H, Yaseri M. Aspheric versus spherical posterior chamber intraocular lenses. J Ophthalmic Vis Res 2010; 5(4): 217-22.
9
Assaf A, Kotb A. Ocular aberrations and visual performance with an aspheric single-piece intraocular lens: Contralateral comparative study. J Cataract Refract Surg 2010; 36(9): 1536-42.
10
Srivannaboon S, Chirapapaisan C. Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery. Graefes Arch Clin Exp Ophthalmol 2019; 257(12): 2677-82.
11
Reinstein DZ, Archer TJ, Srinivasan S, et al. Standard for reporting refractive outcomes of intraocular lens–based refractive surgery. J Refract Surg 2017; 33(4): 218-22.
12
Borkenstein AF, Borkenstein EM. Patient and surgeon satisfaction levels after using an acrylic, hydrophobic, monofocal iol and the malyugin ring in pseudoexfoliation syndrome patients. J Ophthalmol 2018; 2018: 3843098.
13
Kahraman G, Schrittwieser H, Walch M, et al. Anterior and posterior capsular opacification with the Tecnis ZCB00 and AcrySof SA60AT IOLs: a randomised intraindividual comparison. Br J Ophthalmol 2014; 98(7): 905-9.
14
Ong HS, Subash M, Sandhu A, Wilkins MR. Intraocular lens delivery characteristics of the preloaded AcrySof IQ SN60WS/AcrySert injectable lens system. Am J Ophthalmol 2013; 156(1): 77-81.e2.
15
Avadhani K, Shakunthala A, Thirumalai SM, Prakash G. Tale of a twist: Progressive postoperative intraocular lens tilt from a twisted haptic. J Cataract Refract Surg 2015; 41(5): 1111-3.
16
van Vreeswijk H. Safe and easy way to release sticking haptic of a single-piece AcrySof intraocular lens. J Cataract Refract Surg 2008; 34(9): 1611.
17
Shah U. Probable edge defect in Acrysof single-piece intraocular lens. Indian J Ophthalmol 2006; 54(4): 277-8.
18
Jung GB, Jin KH, Park HK. Physicochemical and surface properties of acrylic intraocular lenses and their clinical significance. J Pharm Investig 2017; 47(5): 453-60.
19
Chung B, Lee H, Choi M, Seo KY, Kim EK, Kim TI. Preloaded and non-preloaded intraocular lens delivery system and characteristics: human and porcine eyes trial. Int J Ophthalmol 2018; 11(1): 6-11.
20
Bellucci R, Morselli S, Pucci V. Spherical aberration and coma with an aspherical and a spherical intraocular lens in normal age-matched eyes. J Cataract Refract Surg 2007; 33(2): 203-9.
21
Denoyer A, Le Lez ML, Majzoub S, Pisella PJ. Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation. J Cataract Refract Surg 2007; 33(2): 210-6.
22
Krall EM, Arlt EM, Jell G, Strohmaier C, Moussa S, Dexl AK. Prospective Randomized Intraindividual Comparison of Posterior Capsule Opacification After Implantation of an IOL With and Without Heparin Surface Modification. J Refract Surg 2015; 31(7): 466-72.